MEETINGS / EVENTS

RSS

January 23 - 25 2015, 12:00 AM - 12:00 AM

ASA PRACTICE MANAGEMENT 2015

February 07 - 08 2015, 12:00 AM - 12:00 AM

ASA Certificate in Business Administration 2015

June 26 - 28 2015, 12:00 AM - 12:00 AM

Annual Perioperative Surgical Home Summit

>

FDA MEDWATCH ALERTS

November 07, 2014

FDA MedWatch - Nellcor Puritan Bennett, 980 Ventilator System: Class I Recall - Software Issue May Stop Ventilator

Summary:

FDA MedWatch Nellcor Puritan Bennett 980 Ventilator System Class I Recall

October 25, 2014

FDA MedWatch - Intravia Containers by Baxter: Recall - Particulate Matter

Summary:

FDA MedWatch Intravia Containers by Baxter Recall

October 20, 2014

Lidocaine HCI Injection, USP 10 MG Per ML, 30 ML Single-Dose, Preservative-Free, by Hospira: Recall - Particulate Matter

Summary:

FDA Medwatch Alert 10-20-14

>

ASA FEATURED PRODUCT

Add this product to your shopping cart

Self-Education and Evaluation (SEE) Program

SKU: 30701-14CE

... Read more »

Single Copies, Member Price: $360

Cardiac Science Powerheart, CardioVive, CardioLife; GE Responder and Responder Pro; and Nihon-Kohden Automated External Defibrillators (AEDs): Class I Recall - Defective Component

Monday, March 05, 2012

Affected Models include:

  • Powerheart 9300A, 9300E, 9300P, 9390A, and 9390E
  • CardioVive 92532, 92533
  • CardioLife 9200G and 9231
  • GE Responder and Responder Pro
  • Nihon-Kohden AEDs 

AUDIENCE: Emergency Medicine, Cardiology, Risk Manager

ISSUE: FDA notified healthcare professionals and medical care organizations of the Class 1 recall of the listed AEDs which contain a component that may fail unexpectedly due to a defect. If the component were to fail during a rescue attempt, the AED may not deliver defibrillation therapy, causing serious adverse health consequences, including death. The unit’s self test may not detect the failure or impending failure of the component.

BACKGROUND: These products are used for emergency treatment of victims showing symptoms of sudden cardiac arrest who are unresponsive and not breathing. These AEDs were manufactured and distributed from July 1, 2011 through December 30, 2011.

RECOMMENDATION: Affected customers are advised to contact the firm to arrange for delivery of shipping materials for an immediate return of their AEDs for repair. The affected devices will receive a hardware correction, and the same serial number device will be returned to the customer in most cases.

Read the MedWatch safety alert, including links to the Recal Notice and Letter, at:

http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm294538.htm