May 16, 2012

Hydromorphone Hydrochloride Recall

Summary:

Hydromorphone Hydrochloride Recall

April 18, 2012

Morphine Sulfate Injection USP, 4 mg/mL (C-II), 1 mL fill in 2.5 mL Carpuject by Hospira, Inc: Recall - May Contain More Than Intended Fill Volume

Summary:

Customer report of two Carpujects syringes containing more than the 1 mL labeled fill volume. Opioid pain medications such as morphine have life-threatening consequences if overdosed. Those consequences can include respiratory depression (slowed breathing or suspension of breathing), and low blood pressure.

March 05, 2012

Cardiac Science Powerheart, CardioVive, CardioLife; GE Responder and Responder Pro; and Nihon-Kohden Automated External Defibrillators (AEDs): Class I Recall - Defective Component

Summary: FDA notified healthcare professionals and medical care organizations of the Class 1 recall of the listed AEDs which contain a component that may fail unexpectedly due to a defect. If the component were to fail during a rescue attempt, the AED may not deliver defibrillation therapy, causing serious adverse health consequences, including death. The unit’s self test may not detect the failure or impending failure of the component.

>

ASA FEATURED PRODUCT

Anesthesiology Continuing Education (ACE) Program

SKU: 30702-12CE

... Read more »

Single Copies, Member Price: $300

Oridion Medical and Philips Healthcare Microstream CO2 Filterline (FilterLine H Set Infant/Neonate, VitaLine H Set Infant/Neonate: Recall - Plastic Strands on the Adapter May Become Dislodged, Inhaled by Patient

Tuesday, June 07, 2011

ISSUE : Fine plastic strands on the inner surface of the infant/neonatal airway adapter may become dislodged and inhaled by the patient. Inhalation of the plastic strands on the defective devices may cause respiratory compromise, which could result in serious illness or death.

BACKGROUND : The Microstream CO2 Filterline is a medical device used by emergency medical services, hospitals, and other health care providers to measure exhaled carbon dioxide (CO2) during ventilation of newborn and infant patients.  See the recall notice for affected models and lots.

RECOMMENDATION : Philips instructed users to immediately identify all products from affected lots, remove them from inventory and dispose in accordance with local regulations.
Health care professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety

Information and Adverse Event Reporting Program:

- Complete and submit the report online: www.fda.gov/MedWatch/report.htm.
- Download the form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178.

Read the MedWatch safety alert, including links to the Class I Recall Notice and press release, at: http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm258133.htm