The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.
This recall is not associated with the PE-PUR foam issue impacting certain BiPAP machines recalled in June 2021, described in the Update: Certain Philips Respironics Ventilators, BiPAP Machines, and CPAP Machines Recalled Due to Potential Health Risks: FDA Safety Communication. However, this new recall does apply to some of the devices recalled in June 2021.
A bi-level positive airway pressure (also known as Bilevel PAP, BiPAP, or BPAP) machine pumps air under varying pressure into the airway of the lungs. BiPAP (Ventilator) machines use a higher pressure when you breathe in and lower pressure when you breathe out. These devices are prescribed for people with obstructive sleep apnea syndrome to keep their airways open during sleep, for those who have respiratory insufficiency (not enough oxygen breathed in or carbon dioxide breathed out), or respiratory failure (stop breathing). BiPAP machines may be used to treat adult and pediatric patients at home and in clinical environments, such as hospitals and sleep laboratories, depending on the instructions for use for the particular device model.
Philips Respironics (Philips) is recalling certain BiPAP machines that may contain a plastic contaminated with a non-compatible material. If that plastic is in the device motor, it may release certain chemicals of concern called volatile organic compounds (VOCs). The plastic may also cause the machine to fail and stop working suddenly during use.
The potential risks of inhaling VOCs include:
If the plastic causes the machine to fail and stop working suddenly, it may also lead to serious injury or death.
The FDA is not aware of any reports of serious injury or death associated with the recalled BiPAP machines due to the contaminated plastic issue.
Phillips sent affected customers an Urgent Medical Device Recall letter on August 26, 2022. The contacts included Durable Medical Equipment (DME) suppliers. The letter offered the following recommendations.
For health care providers:
Locate all recalled devices.
For patients and caregivers:
The FDA also issued a safety communication, Certain Philips Respironics BiPAP Machines Recalled Due to a Plastic Issue, for health care providers, patients and their caregivers.
Customers who need any further information or support should contact Philips Recall Support at 1-877-387-3311 or email at [email protected].
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program using an online form, regular mail, or FAX.